da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon Console (SSC), Patient Cart (PSCART), Vision System (VSS)

Intuitive Surgical, Inc. · 20 Sept 2022

Status
Recall open, classified
Why, in the FDA's words
“Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3) Yellow pedal associated with force bipolar is pressed, released with head in surgeon console 4) Blue pedal associated with bipolar is pressed, resulting in force bipolar energy delivery”
Recall began
20 Sept 2022
Company
Intuitive Surgical, Inc.
Device type
Robot-assisted surgical systems (computer-controlled instruments) (NAY)
Concerns
K131861, K171294

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.