Recall

Brand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/A Product Description: Preface: This report under 21 CFR 806.10 is being submitted for the dV5 system tower, as the NIR Handheld…

Intuitive Surgical, Inc. · 24 April 2026

Status
Recall open, classified
Why, in the FDA's words
“Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the handheld camera to the robotic endoscope.”
Recall began
24 April 2026
Company
Intuitive Surgical, Inc.
Device type
Robot-assisted surgical systems (computer-controlled instruments) (NAY)
Concerns
K232610

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.