Recall

da Vinici X da Vinci Xi Sureform 45 REF 480445 da Vinci X da Vinci Xi Sureform 45 Curved-Tip REF 480545 da Vinici X da Vinci Xi Sureform 60 REF 480460

Intuitive Surgical, Inc. · 17 March 2022

Status
Recall open, classified
Why, in the FDA's words
“As a result of complaint data review, it identified tissue pushout events occurring during firing of the SureForm stapler when tissue is pushed distally within the jaws, rather than being held in place and transected. Overall rate of harm associated with bleeding of critical severity exceeds the predicted likelihood of harm as documented within Intuitive s existing risk documentation.”
Recall began
17 March 2022
Company
Intuitive Surgical, Inc.
Device type
Robot-assisted surgical systems (computer-controlled instruments) (NAY)

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.