Monday, October 26, 2026 · Daily editionET --:--:--

Recall

da Vinci Xi and X 12-8 mm Conductive Cannula Reducer, REF: 470381-11

Intuitive Surgical, Inc. · 23 Dec 2021

Status
Recall open, classified
Why, in the FDA's words
“Conductive Cannula Reducer metal tip may get dislodged from the plastic shaft. If unnoticed, additional surgery may be required to remove the metal tip. If the procedure proceeds without the metal tip, there is potential for energy leakage, which could cause tissue damage. If the metal tip remains in the patient, adhesions may form, which may lead to bowel obstruction and chronic abdominal pain.”
Recall began
23 Dec 2021
Company
Intuitive Surgical, Inc.
Device type
Robot-assisted surgical systems (computer-controlled instruments) (NAY)
Concerns
K162411, K140553

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.

The record

The FDA publishes this record in its public device database. The summary documents the company filed, when available, are linked from the FDA's page.
FDA database