Monday, October 26, 2026 · Daily editionET --:--:--

Recall

da Vinci SP Camera; PN 430060-33; Model Number: 430060-33. For endoscopic surgery.

Intuitive Surgical, Inc. · 19 Aug 2019

Status
Recall open, classified
Why, in the FDA's words
“Some cameras were found to have residual soil adhered to the tip of the camera after reprocessing which may affect image quality resulting in needing to switch to a second camera or converting to another surgical modality. If residual soil makes contact with the patient, exposure could result in sensitization making future component blood therapy or tissue typing more challenging; additionally if a small piece of inert epoxy were to be retained, scar tissue could form.”
Recall began
19 Aug 2019
Company
Intuitive Surgical, Inc.
Device type
Robot-assisted surgical systems (computer-controlled instruments) (NAY)
Concerns
K173906

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.

The record

The FDA publishes this record in its public device database. The summary documents the company filed, when available, are linked from the FDA's page.
FDA database