Recall

Da Vinci Xi/X EndoWrist Monopolar Curved scissors (MCS), model number 470179-19, UDI: (00) 886874112298

Intuitive Surgical, Inc. · 4 Dec 2018

Status
Recall terminated
Why, in the FDA's words
“This recall is being initiated because lots of EndoWrist Monopolar Curved Scissor instruments are not recognized by the integrated electrosurgical unit, which prohibits the activation of monopolar energy from the surgeon console.”
Recall began
4 Dec 2018
Company
Intuitive Surgical, Inc.
Device type
Robot-assisted surgical systems (computer-controlled instruments) (NAY)
Concerns
K131861

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.

The record

The FDA publishes this record in its public device database. The summary documents the company filed, when available, are linked from the FDA's page.