EndoWrist Suction Irrigator, Single Use Instrument. Part Number: 480299-04. An accessory to the EndoWrist Surgical System, used in endoscopic surgery.
Intuitive Surgical, Inc. · 4 June 2018
- Status
- Recall open, classified
- Why, in the FDA's words
- “The firm has identified the potential for a silicone particle to be generated within the Suction Irrigator device. If generated, a particle may be introduced into the patient with irrigation.”
- Recall began
- 4 June 2018
- Company
- Intuitive Surgical, Inc.
- Device type
- Robot-assisted surgical systems (computer-controlled instruments) (NAY)
- Concerns
- K162973