Recall
Assembly, Patient Side Manipulator on da Vinci Surgical Systems: da Vinci S Surgical System; da Vinci SI Surgical System da Vinci Si-e Surgical System; Manufactured by Intuitive Surgical Sunnyvale, CA 94086. The Intuitive Surgical…
Intuitive Surgical, Inc. · 11 Nov 2013
- Status
- Recall terminated
- Why, in the FDA's words
- “Reports of friction within certain instrument arms can interrupt smooth instrument motion. This can be felt by the surgeon as resistance in the movement of the master.In this situation, the instrument can stall momentarily and then suddenly catch-up to the master position if the surgeon pushes through the resistance.”
- Recall began
- 11 Nov 2013
- Company
- Intuitive Surgical, Inc.
- Device type
- Robot-assisted surgical systems (computer-controlled instruments) (NAY)