8 mm Monopolar Curved Scissors (a.k.a. Hot Shears) Manufactured and Distributed by: Intuitive Surgical Sunny vale, CA The EndoWrist Monopolar Curved Scissors Instrument is a multiple-use endoscopic instrument utilizing a single use, tip…

Intuitive Surgical, Inc. · 8 May 2013

Status
Recall terminated
Why, in the FDA's words
“Intuitive Surgical has identified a potential for some units of the Monopolar curved Scissors version 09 and 10 instruments to develop very small cracks near the distal end. This may create a pathway for electrosurgical energy to leak to tissue and potentially cause thermal injury.”
Recall began
8 May 2013
Company
Intuitive Surgical, Inc.
Device type
Robot-assisted surgical systems (computer-controlled instruments) (NAY)
Concerns
K050005

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.