CyberKnife Robotic Radiosurgery System with 4D Planning procedure option of MultiPlan Treatment Planning System, versions 3.0, 3.1, 3.5.1, 3.5.2 and 3.5.3 with Ray-Tracing dose calculation. Accuray Incorporated, Sunnyvale, CA. Treatment…

Accuray Incorporated · 16 Jan 2012

Status
Recall terminated
Why, in the FDA's words
“An anomaly was discovered during internal regression testing, where the dose information is displayed incorrectly during treatment planning during a specific workflow using the optional 4D Ray Tracing dose calculation algorithm. As a result, it is possible that the dose calculation will display a lower dose than the intended dose prescribed for treatment delivery.”
Recall began
16 Jan 2012
Company
Accuray Incorporated
Device type
Product code (IYE)
Concerns
K091999

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.