CyberKnife Treatment Delivery System, REF 0660000 The CyberKnife System may be labeled as CyberKnife M6TM or VSI TM Robotic Surgery System. The CyberKnife Treatment Delivery System is indicated for image-guided stereotactic radiosurgery…
Accuray Incorporated · 23 Dec 2024
- Status
- Recall completed
- Why, in the FDA's words
- “Snap rings may become partially or completed detached from the shafts within the internal mechanism. Uncontrolled rotation may occur on the left and right roll axes when the rings are removed or damaged.”
- Recall began
- 23 Dec 2024
- Company
- Accuray Incorporated
- Device type
- Product code (IYE)
- Concerns
- K170788
What it means
A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.