Recall
CyberKnife Treatment Planning System, a subsystem of the CyberKnife Robotic Radiosurgery System, with MultiPlan Treatment Planning System Software version 3.5 Medical charged-particle radiation therapy system, intended for treatment…
Accuray Inc · 29 Dec 2009
- Status
- Recall terminated
- Why, in the FDA's words
- “If electron density values are left empty, the calculation of radiation dose in a patient will be modeled as air-like density material rather than correct density. A plan may be created and saved, thus creating the risk of mistreatment.”
- Recall began
- 29 Dec 2009
- Company
- Accuray Inc
- Device type
- Product code (MUJ)
- Concerns
- K091146
What it means
A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.