CyberKnife System Robotic Radiosurgery System, (Standard Treatment Couch) Synchrony Respiratory Tracking System, a Radiation Therapy Device, Model Numbers: 026429, 028185, Product is manufactured and distributed by Accuray, Inc.,…

Accuray Inc · 16 March 2009

Status
Recall terminated
Why, in the FDA's words
“The product has the potential for the extension coupling of the Synchrony extension mounting hardware to become loose over time. The assembly weighs approximately 17 pounds; an unexpected contact with users or patients may result in serious injury if the boom descends unexpectedly.”
Recall began
16 March 2009
Company
Accuray Inc
Device type
Product code (IYE)
Concerns
K072504

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.