INTUITIVE SURGICAL DA VINCI SURGICAL SYSTEM, MODE; S Cardiac Probe Gasper Instrument, Model IS2000, Part Number 420215-02; Product is manufactured and distributed by Intuitive Surgical, Inc. Sunnyvale, CA . The IS2000 CPG is used during…

Intuitive Surgical, Inc. · 18 April 2008

Status
Recall terminated
Why, in the FDA's words
“The product has a software interface problem. When the product is connected to the da Vinci IS2000 System it will not allow the System to recognize the Instrument which makes the IS2000 CPG instrument nonfunctional at all sites. Risks associated are loss of operability of the instrument, delay in surgery, and loss of dexterity.”
Recall began
18 April 2008
Company
Intuitive Surgical, Inc.
Device type
Robot-assisted surgical systems (computer-controlled instruments) (NAY)
Concerns
K070947

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.