Recall
Intuitive brand da Vinci S Surgical System, Endoscopic Instrument Control System, Model IS1200 with Isolation Transformer (P/N: 952012) and Vision Cart (Model VS1000), Distributed and Manufactured by: Intuitive Surgical, Inc. 950 Kifer…
Intuitive Surgical, Inc. · 4 Oct 2007
- Status
- Recall terminated
- Why, in the FDA's words
- “Under-Rated Fuses--The Firm may have installed incorrect fuses. The daVinci Surgical System Vision Cart's isolation transformer requires a 10A fuse for 115V operation or a 5A fuse for 230V operation. Operation of this product with the use of an under-rated fuse will result in failure of the fuse and a loss of power to the vision cart, including the loss of power to any ancillary equipment that m”
- Recall began
- 4 Oct 2007
- Company
- Intuitive Surgical, Inc.
- Device type
- Robot-assisted surgical systems (computer-controlled instruments) (NAY)
What it means
A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.
The record
The FDA publishes this record in its public device database. The summary documents the company filed, when available, are linked from the FDA's page.
