Recall
Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System
Orthalign, Inc · 10 Feb 2026
- Status
- Recall open, classified
- Why, in the FDA's words
- “Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)”
- Recall began
- 10 Feb 2026
- Company
- Orthalign, Inc
- Device type
- Orthopedic stereotaxic instruments (knee, hip and spine robots) (OLO)
- Concerns
- K200892
What it means
A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.
The record
The FDA publishes this record in its public device database. The summary documents the company filed, when available, are linked from the FDA's page.