Recall
Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Navigation Straight Bone Probe, REF: PC2-400013; Navigation, 3.5mm Drill, REF: PC2-400135; Navigation, 4.0mm Drill, REF:…
SEASPINE ORTHOPEDICS CORPORATION · 17 Nov 2025
- Status
- Recall open, classified
- Why, in the FDA's words
- “The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments. However, the NorthStar OCT Navigation Instruments are compatible for use with Medtronic s Solera toolcards only.”
- Recall began
- 17 Nov 2025
- Company
- SEASPINE ORTHOPEDICS CORPORATION
- Device type
- Orthopedic stereotaxic instruments (knee, hip and spine robots) (OLO)
- Concerns
- K172517
What it means
A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.
The record
The FDA publishes this record in its public device database. The summary documents the company filed, when available, are linked from the FDA's page.
