Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Navigation Straight Bone Probe, REF: PC2-400013; Navigation, 3.5mm Drill, REF: PC2-400135; Navigation, 4.0mm Drill, REF:…

SEASPINE ORTHOPEDICS CORPORATION · 17 Nov 2025

Status
Recall open, classified
Why, in the FDA's words
“The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments. However, the NorthStar OCT Navigation Instruments are compatible for use with Medtronic s Solera toolcards only.”
Recall began
17 Nov 2025
Company
SEASPINE ORTHOPEDICS CORPORATION
Device type
Orthopedic stereotaxic instruments (knee, hip and spine robots) (OLO)
Concerns
K172517

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.