Recall
DYNJ909119G ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909127F CYSTO DYNJ909129D GENERAL LAPAROSCOPY-LF
Medline Industries, LP · 10 Oct 2025
- Status
- Recall open, classified
- Why, in the FDA's words
- “Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.”
- Recall began
- 10 Oct 2025
- Company
- Medline Industries, LP
- Device type
- Product code (FDE)
What it means
A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.
The record
The FDA publishes this record in its public device database. The summary documents the company filed, when available, are linked from the FDA's page.