Monday, October 26, 2026 · Daily editionET --:--:--

Recall

Basic Biopsy Tray; Item Number/Description: AKRB26D ROBOTIC, AMGL09AA GENERAL LAPAROSCOPY, ANLC35AD LAP CHOLE PACK - 205956, ANLP25AJ LAPROSCOPY PACK - 205958, FYLA38E LAPCHOLAPPY PACK, IHDA88AH DIVINCI GYN URO, IHLC11AE LAP GENERAL…

American Contract Systems Inc. · 4 Sept 2025

Status
Recall open, classified
Why, in the FDA's words
“Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.”
Recall began
4 Sept 2025
Company
American Contract Systems Inc.
Device type
Product code (FDE)

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.

The record

The FDA publishes this record in its public device database. The summary documents the company filed, when available, are linked from the FDA's page.
FDA database