StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation System, with StealthStation Spine software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous…
Medtronic Navigation, Inc. · 22 Aug 2024
- Status
- Recall open, classified
- Why, in the FDA's words
- “Due to two software anomalies that can occur within a spine procedure. Two issues are: 1. Navigation Orientation - when lateral orientation radiographic images are taken, information may appear flipped when utilizing the system (180 degree flipped). 2. Pop-up Message - when using CT+Fluouroscopy images with a slice spacing thickness greater than 2mm, the system does not display a pop-up message indicating the slice spacing is not optimal.”
- Recall began
- 22 Aug 2024
- Company
- Medtronic Navigation, Inc.
- Device type
- Orthopedic stereotaxic instruments (knee, hip and spine robots) (OLO)
- Concerns
- K201189
What it means
A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.