Monday, October 26, 2026 · Daily editionET --:--:--

Recall

Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic System

Think Surgical, Inc. · 12 Feb 2024

Status
Recall open, classified
Why, in the FDA's words
“There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling.”
Recall began
12 Feb 2024
Company
Think Surgical, Inc.
Device type
Orthopedic stereotaxic instruments (knee, hip and spine robots) (OLO)
Concerns
K230202

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.

The record

The FDA publishes this record in its public device database. The summary documents the company filed, when available, are linked from the FDA's page.
FDA database