Monday, October 26, 2026 · Daily editionET --:--:--

Recall

ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844

Globus Medical, Inc. · 15 Dec 2023

Status
Recall open, classified
Why, in the FDA's words
“Devices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an increased risk for infection, inflammation, local toxicity, systemic toxicity, sensitivity, and cross-contamination.”
Recall began
15 Dec 2023
Company
Globus Medical, Inc.
Device type
Orthopedic stereotaxic instruments (knee, hip and spine robots) (OLO)
Concerns
K171651

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.

The record

The FDA publishes this record in its public device database. The summary documents the company filed, when available, are linked from the FDA's page.
FDA database