Recall

ExcelsiusGPS Flat Panel Fluoroscopy Fixture Kit, Part number 6203.2100

Globus Medical, Inc. · 7 Dec 2023

Status
Recall open, classified
Why, in the FDA's words
“A calibration error associated with the Flat Panel Fluoroscopy Fixture when used with ExcelsiusGPS for fluoroscopy workflow may lead to inaccurate navigation, which may result in device misplacement.”
Recall began
7 Dec 2023
Company
Globus Medical, Inc.
Device type
Orthopedic stereotaxic instruments (knee, hip and spine robots) (OLO)
Concerns
K211616

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.