ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in Radiation Therapy, Radiology, or Radiology image-guided treatment environment. Part Number: 1SOPT00003

LEONI CIA CABLE SYSTEMS · 6 Oct 2023

Status
Recall open, classified
Why, in the FDA's words
“Users of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the movements are again allowed after an unforeseen interruption signal of motion may result in a collision of the patient with another stationary component of the treatment room.”
Recall began
6 Oct 2023
Company
LEONI CIA CABLE SYSTEMS
Device type
Product code (JAI)
Concerns
K160518

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.