Recall
Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299
Blue Belt Technologies, Inc · 7 Sept 2023
- Status
- Recall open, classified
- Why, in the FDA's words
- “CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational reference via the posterior condylar axis”
- Recall began
- 7 Sept 2023
- Company
- Blue Belt Technologies, Inc
- Device type
- Orthopedic stereotaxic instruments (knee, hip and spine robots) (OLO)
- Concerns
- K231963