Recall

Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299

Blue Belt Technologies, Inc · 7 Sept 2023

Status
Recall open, classified
Why, in the FDA's words
“CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational reference via the posterior condylar axis”
Recall began
7 Sept 2023
Company
Blue Belt Technologies, Inc
Device type
Orthopedic stereotaxic instruments (knee, hip and spine robots) (OLO)
Concerns
K231963

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.

The record

The FDA publishes this record in its public device database. The summary documents the company filed, when available, are linked from the FDA's page.