Magellan Robotic System; Model No(s). 11132 and 11139. A steerable catheter control system used at clinical sites.

Hansen Medical Inc · 18 March 2015

Status
Recall terminated
Why, in the FDA's words
“Based on investigation of two complaint incidents, all Hansen Medical Magellan Robotic System (Model No. 11132 and 11139) may experience an unexpected event during a retraction operation of catheter. Retraction of catheter may not stop when equipment button is released.”
Recall began
18 March 2015
Company
Hansen Medical Inc
Device type
Product code (DXX)
Concerns
K132369, K141614

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.