Monday, October 26, 2026 · Daily editionET --:--:--

Recall

ACCELERATOR APS Centrifuge Module (Hettich Robotic centrifuge); List Number: 07L02-01; manufactured by Inpecco SpA, Segrate (MI), Italy.

Abbott Laboratories, Inc · 9 Sept 2008

Status
Recall terminated
Why, in the FDA's words
“ACCELERATOR APS Centrifuge Module being recalled due to potential for micro-cracks to form over time in the metal centrifuge rotor buckets that may cause the centrifuge buckets to disconnect from the rotor in the centrifuge during centrifugation.”
Recall began
9 Sept 2008
Company
Abbott Laboratories, Inc
Device type
Product code (JQP)

What it means

A recall is the FDA's record that a company is correcting or removing a device because of a problem. Many are fixed with a software update or new instructions; the record does not say whether anyone was hurt. See our plain guide to FDA terms.

The record

The FDA publishes this record in its public device database. The summary documents the company filed, when available, are linked from the FDA's page.
FDA database